Abstract

A new parallel-plate capacitor fixture has been designed and successfully used to measure dielectric loss of polyelectrolyte solutions with volumes as low as droplets of 13–26 μL. It is particularly useful when studying polypeptides that are either high-cost or can be synthesized only in limited quantities. The ease with which the fixture can be used to obtain preliminary dielectric loss data yields savings in time and cost. In this study capacitance measurements were performed in a wide range of frequencies between 1 and 800 MHz using an Agilent 4191RF Impedance Analyzer. Accuracy of measurements was carefully examined through a comparison of measured conductivity of 1M NaCl against Stogryn's equation for conductivity. A 0.3% difference between the experimentally measured and theoretically calculated results has been found, demonstrating the validity of the proposed analysis method.

Rights Information

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.

Was this content written or created while at USF?

Yes

Citation / Publisher Attribution

Journal of Science: Advanced Materials and Devices, v. 1, issue 4, p. 521-526