Intrinsic Dependencies of CT Radiomic Features on Voxel Size and Number of Gray Levels
Document Type
Article
Publication Date
2017
Keywords
computed tomography, features, gray levels, phantom, radiomics, texture, voxel size
Digital Object Identifier (DOI)
https://doi.org/10.1002/mp.12123
Abstract
Purpose: Many radiomics features were originally developed for non-medical imaging applications and therefore original assumptions may need to be reexamined. In this study, we investigated the impact of slice thickness and pixel spacing (or pixel size) on radiomics features extracted from Computed Tomography (CT) phantom images acquired with different scanners as well as different acquisition and reconstruction parameters. The dependence of CT texture features on gray-level discretization was also evaluated.
Methods and materials: A texture phantom composed of 10 different cartridges of different materials was scanned on eight different CT scanners from three different manufacturers. The images were reconstructed for various slice thicknesses. For each slice thickness, the reconstruction Field Of View (FOV) was varied to render pixel sizes ranging from 0.39 to 0.98 mm. A fixed spherical region of interest (ROI) was contoured on the images of the shredded rubber cartridge and the 3D printed, 20% fill, acrylonitrile butadiene styrene plastic cartridge (ABS20) for all phantom imaging sets. Radiomic features were extracted from the ROIs using an in-house program. Features categories were: shape (10), intensity (16), GLCM (24), GLZSM (11), GLRLM (11), and NGTDM (5), fractal dimensions (8) and first-order wavelets (128), for a total of 213 features. Voxel-size resampling was performed to investigate the usefulness of extracting features using a suitably chosen voxel size. Acquired phantom image sets were resampled to a voxel size of 1 × 1 × 2 mm3using linear interpolation. Image features were therefore extracted from resampled and original datasets and the absolute value of the percent coefficient of variation (%COV) for each feature was calculated. Based on the %COV values, features were classified in 3 groups: (1) features with large variations before and after resampling (%COV >50); (2) features with diminished variation (%COV
Results: Out of the 213 features extracted, 150 were reproducible across voxel sizes, 42 improved significantly (%COV
Conclusion: Voxel-size resampling is an appropriate pre-processing step for image datasets acquired with variable voxel sizes to obtain more reproducible CT features. We found that some of the radiomics features were voxel size and gray-level discretization-dependent. The introduction of normalizing factors in their definitions greatly reduced or removed these dependencies.
Was this content written or created while at USF?
Yes
Citation / Publisher Attribution
Medical Physics, v. 44, issue 3, p. 1050-1062
Scholar Commons Citation
Shafiq-ul-Hassan, Muhammad; Zhang, Geoffrey G.; Latifi, Kujtim; Ullah, Ghamim; Hunt, Dylan C.; Balagurunathan, Yoganand; Abdalah, Mahmoud Abrahem; Schabath, Matthew B.; Goldgof, Dmitry G.; Mackin, Dennis; Court, Laurence Edward; Gillies, Robert James; and Moros, Eduardo Gerardo, "Intrinsic Dependencies of CT Radiomic Features on Voxel Size and Number of Gray Levels" (2017). Physics Faculty Publications. 127.